Working with alarms
When an alarm is muted, the Alarm mute key LED will light constantly. If after two minutes the
problem is still present, the alarm will sound again for high or medium priority alarms except for the
Internal Battery Low (30%) alarm. This alarm will be permanently muted with both the audible signal
and alarm message cleared. However, these indicators will initiate again when the Internal Battery
Empty (2 minutes remaining) alarm is activated.
Technical specifications
Alarm Volume Range
Applied parts
(as per IEC 60601-1)
FiO
measurement
2
SpO
measurement
2
Internal battery
* The measured FiO
sampled at a minimum of 1Hz has a response time of 14s to 90% of the final value. The long
2
term output drift in 100% O
General warnings and cautions
WARNING
Single-use products are not to be re-used due to degradation and contamination risks.
Symbols
Device weight;
2
High, medium and low priority alarm (measured according to IEC 60601-1-8: 2012):
minimum 42 dB
maximum 79 dB
Patient interface (Mask, endotracheal tube or tracheostomy tube)
Oximeter (optional accessory)
Range: 18% to 100%, Resolution: 1%, Accuracy: ± (2.5 vol.% +2.5% of actual oxygen concentration)*
(as per ISO 80601-2-55)
Range: 70% to 100%, Resolution: 1%, Accuracy: ±2% in a range of 70% to 100%
Lithium-Ion battery, 14.4 V, 2.75 Ah, 40 Wh
Operating hours: 3 h with a new battery under normal conditions
is typically <5% over 1 year.
2
Contains no China environmental hazardous substances.