Start Up ..........................4-2 Changing the Set Point ......................4-3 Selecting a Preset Set Point ....................4-3 Changing a Setting ......................4-4 Selecting a Preset ........................ 4-6 Changing a Preset ....................... 4-6 Temperature Display ......................4-7 Shut Down ........................... 4-7 Thermo Scientific Precision...
Page 4
Precision Water Bath Quick Start Guide This quick start guide is intended for initial start up only. For all other procedures you must refer to the manual. Also, if any of these steps are not clear refer to the manual before proceeding.
Page 5
Changing the Set Point When operating without a lid, limit the maximum set point to 60°C. You can change the set point with the bath running or not. From the home screen press either arrow key to display: Press, and hold, the arrow keys to bring up the desired set point value. The range is 5°C to 100°C.
Page 6
Kurzanleitung zur Inbetriebnahme des Präzisionswasserbades Diese Kurzanleitung ist nur für die erste Inbetriebnahme vorgesehen. Für alle anderen Verfahren müssen Sie im Handbuch nachsehen. Auch wenn irgendwelche Schritte unklar sind, schaue Sie im Handbuch nach, bevor Sie fortfahren. Befüllung Stellen Sie sicher, dass der Ablauf des Behälters geschlossen ist und dass alle Anschlüsse sicher angeschlossen oder verschlossen sind. Achten Sie außerdem darauf, dass vor dem Befüllen des Behälters alle Rückstände gründlich entfernt wurden.
Page 7
Ändern des Sollwertes Beim Betrieb ohne Abdeckung ist der Sollwert auf maximal 60 °C zu begrenzen. Sie können den Sollwert ändern, gleichgültig, ob das Bad läuft oder nicht. Vom Einstiegsbildschirm aus drücken Sie eine der beiden Pfeiltasten zur Anzeige von: Drücken Sie die Pfeiltasten und halten Sie diese gedrückt , bis der gewünschte Sollwert erscheint.
Page 8
Guide de démarrage rapide - Bain de précision Ce guide de démarrage rapide est destiné à la mise en marche initiale uniquement. Pour toute autre procédure, merci de vous référer au manuel. De plus, si l’une de ses étapes ne vous paraît pas claire, téléchargez le manuel avant de commencer.
Page 9
Modification de la température définie Si vous utilisez le bain sans couvercle, définissez la température maximum à 60 °C. Vous pouvez modifier la température définie que le bain soit ou non en cours d’utilisation. À l’écran d’accueil, appuyez sur une touche flèche pour afficher : Appuyez sur les touches flèche et maintenez-les enfoncées jusqu’à...
Page 10
Guía de inicio rápido de bañeras de agua de precisión Esta guía de inicio rápido se ha creado únicamente para el arranque inicial. Para obtener información sobre el resto de procedimientos, debe consultar el manual. Asimismo, en caso de que tuviera dudas sobre alguno de estos pasos, consulte el manual antes de continuar.
Page 11
Modificación del punto de ajuste Cuando se opere sin tapa, limite el punto de ajuste máximo a 60 °C. Puede cambiar el punto de ajuste con la bañera en marcha o parada. En la pantalla de inicio, pulse cualquier tecla de flecha para mostrar: Pulse y mantenga pulsadas las teclas de flecha para que aparezca el valor del punto de ajuste deseado.
Page 12
Banho de água de precisão – Guia de início rápido Este guia de início rápido destina-se apenas ao arranque inicial. Para todos os outros procedimentos, consulte o manual. Além disso, se qualquer uma dessas etapas não estiver clara, consulte o manual antes de prosseguir. Preenchimento Certifique-se de que a porta de drenagem do reservatório esteja fechada e que todas as conexões do encanamento estejam firmemente sondadas ou tampadas.
Page 13
Alterar o Ponto de ajuste Ao operar sem tampa, limite o ponto máximo definido para 60 °C. Você pode alterar o ponto de ajuste com o banho funcionando ou parado. A partir da tela inicial, pressione qualquer tecla de seta para exibir: Pressione e segure as teclas de seta para mostrar o valor de ajuste desejado.
Thermo Fisher Scientific has contracted with one or more recycling/ disposal companies in each EU Member State, dispose of or recycle this product through them. Further information on Thermo Fisher Scientific’s compliance with these Directives is available at: www.thermoscientific.com/WEEERoHS Thermo Scientific Precision...
Monday through Friday (Eastern Time) Fax: (603) 436-8411 service.tc.us@thermofisher.com Feedback We appreciate any feedback you can give us on this manual. Please e-mail us at: tcmanuals@thermofisher.com Please include the manual part number and the revision date listed on the front cover. Precision Thermo Scientific...
"dangerous voltage" within the bath's enclosure. The voltage magnitude is significant enough to constitute a risk of electrical shock. This label indicates the presence of hot surfaces. This label indicates read the manual. Observe all warning labels. Never remove warning labels. Thermo Scientific Precision...
Page 17
Refer service and repairs to a qualified technician. Performance of installation, operation, or maintenance procedures other than the ones described in this manual may result in a hazardous situation and will void the manufacturer's warranty and safety compliance. Precision Thermo Scientific...
Chapter 2 General Information Description All Thermo Scientific Precision Water Baths provide constant temperature to applications placed in the reservoir. All have digital displays, four programmable set point temperatures, acoustic and optical alarms, and offer adjustable high temperature protection. The interior of the bath is constructed of stainless steel and is designed for operation with water.
Thermal beads may be used instead of water but they will degrade the bath's uniformity and stability. Bath depth includes bezel, height does not include lid. Thermo Fisher Scientific takes no responsibility for damages caused by the selection of an unapproved fluids. Thermo Fisher Scientific reserves the right to change specifications without notice. Precision Thermo Scientific...
Page 44
Thermal beads may be used instead of water but they will degrade the bath's uniformity and stability. Bath depth includes bezel, height does not include lid. Thermo Fisher Scientific takes no responsibility for damages caused by the selection of an unapproved fluids. Thermo Fisher Scientific reserves the right to change specifications without notice. Thermo Scientific Precision...
Page 45
Chapter 2 General Information Precision Thermo Scientific...
The bath is designed for continuous operation and for indoor use. Never place the bath in a location where excessive heat, moisture or corrosive CAUTION materials are present. Ventilation No special ventilation clearances are required. Cover Installation Figure 3-1 – GP Bath Cover Installation Thermo Scientific Precision...
Once the cord is connected to the bath, connect the other end to the main power source. The following plug types are available: Bath Volts /Hertz/Phase Plug Type GP (All) 100-115/50-60/1 N5-15 200-230/50-60/1 Country Specific 1. Volts ± 10% 15 Amp Outlet Precision Thermo Scientific...
Chapter 3 Installation Approved Fluids The only approved fluids for all Precision baths are: Filtered/Single Distilled Water De-ionized Water For applications requiring resistivity greater than 1 MΩ-cm or maintaining resistivity levels greater than 1 MΩ-cm please call and speak to an applications engineer for additional information.
Page 49
Another option is heat up the bath for ten minutes then stir the beads with a stir rod. Allow two to five hours to equilibrate. Note Thermal beads will degrade the bath's uniformity and stability. Precision Thermo Scientific...
Page 50
Tap water is not normally recommended because the total ionized solids level may be too high. As an example, tap water in the United States averages 171 ppm (of NaCl). The recommended level for use in a water system is between 0.5 to 5.0 ppm (of NaCl). Thermo Scientific Precision...
Remove the drain hose from its housing and insert the fitting into the drain quick disconnect. • When the flow stops remove the hose and reinsert it back into the housing. Figure 3-3 – GP Bath Drain Hose Thermo Scientific Precision...
Steaming Covers for General Purpose Baths Catalog Description Number 1546230Q 4-Hole Concentric Ring Cover for 10L GP Bath 1546231Q 6-Hole Concentric Ring Cover for 20L and 28L GP Bath Figure 3-4 – GP Bath Concentric Ring Covers Precision Thermo Scientific...
Chapter 4 Operation Chapter 4 Operation Controller The Thermo Scientific Precision Water Baths have digital controllers that display the bath’s reservoir fluid temperature, shaker speed and other bath features. Once the circuit protector on the back of the bath is on, press to start/stop the bath.
The set point is the desired reservoir fluid temperature. 30.0°C • Press to start the bath and display the reservoir fluid temperature. If the fluid temperature is below the set point the heater will start. 29.9°C 30.0°C Precision Thermo Scientific...
T3 44.5°C change and return to the home screen. T4 45.5°C After a 30 second delay if neither key is pressed the display returns to the home screen, any change saved. Thermo Scientific Precision...
. See Chapter 6. Press, and hold, an arrow key to change the temperature. Once the desired value is displayed press to save the change or 95. °C not save the change. Press to return to the home screen. Precision Thermo Scientific...
Page 58
This mode also derates the bath’s stability and uniformity capability. Press to return to the home screen. Thermo Scientific Precision...
RTA 0.0°C and hold, the arrow keys to change the value. Once the desired value is displayed press If desired, press the arrow keys to highlight another preset parameter or press to return to the home screen. Precision Thermo Scientific...
For bath shut down place the circuit protector located on the rear of the bath to the O position. • • To disconnect the bath remove the electrical plug from its power source. Thermo Scientific Precision...
If corrosion (e.g., small rust marks) occur in spite of this, cleaning with stainless steel caustic agents has proved to be suitable. Apply these substances according to the manufacturer's recommendations. Electrical Power Cord Ensure any replacement cord is properly rated. Thermo Scientific Precision...
HTC fault cannot be cleared, the bath must be serviced by an authorized Thermo Scientific Temperature Control Service Technician. High Temp Fault adjustable high temp fault protection limit exceeded...
Please contact Thermo Fisher Scientific Sales Service and Customer Support if you need any additional information, see inside cover for contact instructions. Precision Thermo Scientific...
The buyer shall be responsible for any evaluation or warranty service call (including labor charges) if no defects are found with the Thermo Scientific product. This warranty does not cover any bath that has been subject to misuse, neglect, or accident. This warranty does...