CLEANING & MAINTENANCE
f. Covers may be immersed and soaked in disinfectant for the required incubation period. After
pre-soaking, the cover may be rinsed through a regular cycle in a washer with no soap then
laundered with mild detergent (wash temperature 93°F/34°C, rinse temperature 78°F/26°C or
on the coldest setting).
g. Allow all covers and parts to aerate until they are fully dry.
5) Repeat the process with the tubing set: spray/wipe, incubate, and air dry.
6) Dry the mattress on a flat surface area after cleaning, away from exposure to the sun.
7) Avoid long exposure to sunlight.
HANDLING AND STORAGE
• Lay the mattress out flat and upside down.
• Roll from the foot end towards the head end; the foot-end strap can then be stretched around the rolled
mattress to prevent unrolling.
• Do not fold, crease, or stack the mattress.
• Place the complete system into the carry bag. Do not store in direct sunlight, extreme high or low
temperatures or moist area.
MAINTENANCE
General
• Check the power cord and plug to see if there are abrasions or excessive wear.
• Check the mattress cover for signs of wear or damage. Ensure the mattress cover and tubes are
connected correctly.
• Plug in the control unit and check the airflow from the hose connection port. The airflow should alternate
between ports every half-cycle time.
• Check the air hoses to see if there are any kinks or breaks. For replacement, please contact your local
agent or dealer.
• Make sure the mattress tube is well connected.
• Check the control unit and make sure both power indicators are off when the switch is turned off.
Low pressure
Examine if there is any air leakage between the control unit and the mattress connections or from the air
mattress tubes:
1) Check connectors between the air mattress and control unit. If there is any disconnection, please
reconnect it.
2) Check the air-connecting tubes. Ensure each single cell is properly functioning.
3) Set the pressure at Auto Firm. Keep the tubes fully inflated and inspect for air leakage.
4) Check if there is any air leakage from cells. Ensure no leakage occurs. If any leakage occurs, please
contact your local agent or dealer.