Capacitance (Acd-23Sw Only); Temperature (Acd-23Sw Only); Continuity - Amprobe ACD-22SW Mode D'emploi

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

40.00Me
Open circuit volts: -0.45V dc typical, (-1.2Vdc on 400e range)
Overload protection: 600VDC or AC rms
Capacitance (ACD-23SW only) (Auto Ranging)
Ranges
4.000μF
40.00μF, 400.0μF
4.000mF
Minimum input range: >100nF
Overload protection: 600VDC or AC rms

Temperature (ACD-23SW only)

Ranges
0.0°C to 400.0°C
-35.0°C to 0.0°C
32.0°F to 400.0°F
-30.0°F to 32.0°F
Sensor type: K-type thermocouple
Overload protection: 30V Max
Diode Test (ACD-22SW only)
Test current: 0.8 mA (approximate)
Accuracy: ± (3.0% rdg + 3 dgts)
Open circuit volts: 3.0 Vdc typical
Audible indication: < 0.25V

Continuity

Range: 400e
Resolution: 1e
Audible indication: Less than 25e
Response time: 500ms
Overload protection: 600VDC or AC rms
± (3.0% rdg + 5 dgts)
Accuracy
±(3.0% rdg + 15 dgts)
±(3.0% rdg + 5 dgts)
±(5.0% rdg + 20 dgts)
Accuracy
± (1.0% + 1°C)
± (2.0% + 3°C)
± (1.0% + 2°F)
± (2.0% + 6°F)
10