11
Cleaning
1) Lens:
The lens can be cleaned by using a moistened cotton-tipped applicator with a solution of water with mild detergent.
Do not use abrasive Ultrasonic cleaners, alcohol or other chemicals to clean the lenses as this may degrade lens
quality, as they will scratch the coated surface of the lenses.
2) Handle:
The handle can be cleaned with a moist cloth. Battery handle shall not be dipped into water. While cleaning, make sure
that lamp and battery are not soaked with water. Remove battery and transformer casing while cleaning. Periodically
check the battery condition, making sure that no sign of corrosion or oxidation is present, in case they are present, it
is highly recommended to use new battery. For battery replacement see "Assembling".
Autoclaving and Flash sterilization the handle will result in severe electrical damages.
Storage
Since the product is made of corrosion-proof materials suitable for the environmental conditions foreseen for its
normal use, does not require special care, however it is necessary to store it in such a way to make sure that it is
protected from dust, and water to assure its hygienic conditions and protection against short circuit. Moreover, it is
recommended to store the product without battery.
Use and Maintenance
Before removing the diagnostic unit (otoscope head or ophthalmoscope head), make sure that switch is off. Handle
plug pins case carefully as it contains circuit for charging battery.
Any unauthorized modification/service is not allowed because it may cause short circuit that may damage the equipment,
so always contact the authorized person for repair.
Disconnect the handle from main supply prior to any maintenance.
Handle is not suitable for use in the presence of a flammable anesthetic mixture with air or oxygen or nitrous oxide.
An explosion may result.
Examine heads, handles and transformer case for wear or damage monthly. After every use, clean the Handle
according to the Cleaning Section of this manual.
Caution: read instructions (warnings)
carefully
Keep in a cool, dry place
Manufacturer
Product code
Medical Device compliant with Regulation
(EU) 2017/745
WEEE disposal
Disposal: The product must not be disposed of along with other domestic waste. The users
must dispose of this equipment by bringing it to a specific recycling point for electric and elec-
tronic equipment.
GIMA WARRANTY TERMS
The Gima 12-month standard B2B warranty applies.
ENGLISH
Follow instructions for use
Keep away from sunlight
Date of manufacture
Lot number
Type BF applied part
Medical Device