Operating Instructions
1. The handle is only to be used by authorized professional personnel. The handle is to be used exclusively for
charging 3.5V Li-ion rechargeable battery and for diagnoses purpose.
2. For charging battery, open the end cap located on the plug pins case. Insert plug pins directly to supply mains.
After connection of the plug in handle with the main power source, the handle is ready for charging.
3. It does not have indication that shows that battery is fully charged. So, it is recommended that do not charge battery
for more than 8 hours. However there is no chance of over charging of batteries because batteries contains over
charging cut-off circuit.
4. To stop charging of battery remove Plug pins from mains supply.
5. Before using it for diagnosis purpose, it is necessary to insert the batteries into the handle compartment as illustrated
in assembling (fig.5).
6. Connection between head and handle is through bayonet locking system. In this case, it is necessary to hold the
battery handle and head in each hand and match the bayonet grooves on both the head and handle. For locking
push the head down and turn the head clock wise.
7. Check that the connection between the head and handle is perfect and that the rheostat works correctly.
8. Push the button on the handle and twist the black knob towards the left hand side. The unit is now in operating
position.
9. The intensity of the light can be adjusted by turning the knob clock wise or counter clock wise.
10. Switch off the unit after use to preserve batteries. Twist the black knob towards right until an audible click sound
is heard once the button comes to its off position.
Specifications
Classification: Class II, Type BF Applied Part
Supply Voltage: 100V-240V AC, 50-60 Hz
Output Voltage 3.5-3.7 V
Charging Current: 200 mA
Power Consumption: 2 VA
Battery:
Voltage: 3.5V
Capacity: 1400 mah
Charging time: 7 hours
Type: Li-ion
Lamp:
Voltage: 3.5V
Power: 2.5watt
Lux: 23,000
Life: 20 Hours
Dimensions:
Handle: 143.7mm (5.56") x 32mm (1.25"),
Weight: 151 grams (5.32 oz)
Environment:
Operating: +10°C to +40°C (50° F to 104° F), 95% RH, 500hPa to 1060hPa
Transport/Storage: -20°C to +49°C (-2°F to 120°F), 95% RH, 500hPa to 1060hPa.
Precautions
The handles shall be handled carefully. The product shall be used by qualified personnel only. Do not use the equipment
in case it is damaged. Apply to
your retailer. Broken bulbs can cause injuries or cut. Carefully handle and immediately replace. Avoid precarious
repairs. Repairs shall be carried out with original spare parts only, which shall be installed according to the intended
use. Protect lamp surface against abrasions and scratches.
This product is designed for intermitted use typical on time continuous should not increase 10 minutes.