I-Tech MIO-SONIC Notice D'utilisation page 44

Masquer les pouces Voir aussi pour MIO-SONIC:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

ULTRASOUND HEAD STATIONARY IN ONE POINT DURING THE
THERAPY.
7.
At the end of the therapy all the lights indicator
will shut off. Move the switch on the OFF button
and unplug the device from its power supply.
8.
Clean the ultrasound head from the gel before
storing the device and its accessories in the bag. DO
NOT IMMERSE THE DEVICE IN WATER!
N.B. before storing the device and its accessories in the bag, disconnect the
cables. If this does not do, the cables may take excessive folds near the
connectors, that can cause a damage of the cables.
ATTENTION: it is recommended to use the ultrasound with the
intensity set on M. for H intensity is recommended maximum
attention in maintaining in continuous moving the ultrasound head.
The intensity L corresponds approx. to 0,08W/cm², the intensity M
approx. to 0,80W/cm² and intensity H approx. to 1,6W/cm².
ATTENTION: to ensure the safety of the patient, the device is
equipped with a recognition system of correct coupling between the
ultrasound head and the patient skin. In case of non-correct
coupling or inadequate contact, the light related to the period of
therapy will start light up.
ATTENTION: to ensure the safety of the patient, the device is also
equipped of a system for regulating the temperarure. In the case the
ultrasound head excees 42°C, the device will stop the treatment
and time indicator light will flash two times; it will not be possible
continue the treatment program until the temperature is below
40°C.
IACER Srl
40
MNPG71-04

Publicité

Table des Matières
loading

Table des Matières