Télécharger Imprimer la page

Laica HTD8813 Instructions Et Garantie page 13

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 43
English
EN
measurement in the ear; the digital thermometer for traditional rectal, axillary and oral measurement.
Depending on the point where the measurement is made, a different temperature value is obtained.
The variation can be between 0.2-1°C. You can not compare the temperatures measured with different
measuring methods. It is therefore necessary to consider in case of self-diagnosis or to report to your doctor
what type of thermometer and in which point of the body the temperature has been measured.
PRODUCT DESCRIPTION (see fig.1)
1) Measurement probe
2) "O/I" key
3) "SET" key
4) "MEMO" key
5) "MODE" key
6) Battery compartment
7) Batteries
8) LCD display
a. Indication of ambient and liquid temperature measurement
b. Indication of forehead temperature measurement
c. Temperature recorded
d. Indication of the measurement result
e. Low battery indicator
f. Sound signal active
g. Unit of measure
h. Memory number
TECHNICAL SPECIFICATIONS
• Product name: infrared thermometer
• Trade name: TH1003
• Classification of installation and use: internal energy, portable, continuous use, applied part of class BF, IP22, no
AP or APG
• Range of body temperature measurement: 34°C to 43°C (93.2°F to 109.4°F)
• Environmental temperature and liquid measuring range: 0°C to 100°C (32°F to 212°F)
• Display resolution: 0.1°C/0.1°F
• Backlit LCD display in 3 colours (for body temperature only):
24
34 – 37.3°C/93.2-99.1°F: green, normal temperature;
37.4-38°C/99.3-100.4°F: yellow, fever;
38.1-43°C/100.6-109.4°F: red, high fever
• Measurement distance: 1 - 5 cm
• Operating mode:
forehead measurement ("body"): regulated mode
"surface" measurement: direct mode
• Nominal output range:
forehead measurement ("body"): 34°C~ 43°C (93.2°F-109.4°F)
"surface" measurement 0°C ~ 100°C (32°F-212°F)
• Batteries: 2x1.5V alkaline AAA
• Switch off: automatic, after 18 seconds of non-use
• Battery life: approximately 3000 readings (turning the thermometer on and off after each measurement)
• Operating conditions: from 15°C to 40°C (59°F at 104°F), RH ≤85%
• Atmospheric pressure: 70 - 106 kPa
• Transportation and storage conditions: from -20°C to 55°C (-4°F at 131°F), RH ≤93%
• Tolerance for body temperature:
from 35 to 42 °C (95-107.6°F) :+/-0.2°C (0.4°F)
from 34°C to 34.9°C (93.2-94.8°F) and from 42.1 to 43°C (107.8-109.4°F): +/- 0.3°C (0.5°F)
• Tolerance for ambient temperature and liquids: +/- 2°C (3.6°F)
• Clinical bias: -0,027°C
• Standard deviation: 0.14°C
• Acceptability limit: 0.26°C
• Clinical repeatability: 0.07°C
INSTRUCTIONS FOR USE
INSERTION/REPLACEMENT OF THE BATTERY
This thermometer operates with 2 x 1.5V AAA alkaline batteries.When first used, and when the display shows
the battery symbol,
, proceed with insertion and/or battery replacement. Open the battery compartment
by sliding the cover in the direction of the arrow, then insert the batteries according to the indicated poles and
close the cover. Dispose of used batteries as indicated in the paragraph "Disposal procedure".
English
25
EN

Publicité

loading