5.11 Description of the remote control commands
.Heater.S
Stop: Switching off of the heater
.E
Error: Message with error number (see page 102)
Format of AutoInfo messages:
(space)!device name"AutoInfo node"
Example: !Oven1".G"
Setup.PowerOn
Simulation of "power on". The method last used is ready for operation.
Setup.Initialize
Setup.Initialize.Select
Setting of default values for the following branches:
param:
Method parameters (sets empty method '
config:
Configuration, branch &Config
setup:
Branch &Setup
assembly:
Branch &Assembly
all:
All values of the entire tree.
Initialization is triggered with &Setup.Initialize $G.
Setup.RamInit
Initialization of the entire working memory. All parameters will be set to de-
fault; error messages will be deleted.
Setup.InstrNo.Value
Device number. Essential for service purposes. The device number must
not be altered.
UserMeth.FreeMemory
$Q queries the size of free memory (in bytes), available for user methods.
UserMeth.Recall
UserMeth.Recall.Name
UserMeth.Store
UserMeth.Store.Name
UserMeth.Delete
UserMeth.Delete.Name
UserMeth.DelAll
Management of the internal method memory: loading, storing and deletion
of methods. The execution of a function is performed by sending $G to the
corresponding node, after entering the method name. Do not use preced-
ing or succeeding blanks in method names!
.DelAll;$G deletes all method in the user memory.
UserMeth.List.1.Name
UserMeth.List.Bytes
for each method
List of all method in the user method memory including method name and
file size in bytes.
$G
$G
param, config, setup,
assembly, all
∗∗∗∗∗∗∗∗
')
$G
8 ASCII characters
read only
$G
8 ASCII characters
$G
8 ASCII characters
$G
8 ASCII characters
$G
read only
read only