Instructions On Using Usb Memory Devices; Measurment Datasets; Lement Of A Measurement Dataset; Operations With Measurement Datasets - Secomam UviLine Serie Manuel D'utilisation

Spectrophotometre
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Measurement data
Concentration
Absorption / % transmission
Multi-wavelengths
Spectrum
Kinetics
As the location for the PC readable files (*.csv) to be stored, it is possible to select for each export either the
spectrophotometer (Internal DataB folder) or an external memory (USB memory).
Later it is possible to transmit the files stored in the photometer (Internal DataB folder) to a connected PC or an external
memory (USB memory).
It is possible to transmit to a USB memory either individual files with measuring data or all files stored in the Internal
DataB folder.
1.54

INSTRUCTIONS ON USING USB MEMORY DEVICES

The safety of data stored on USB memory devices depends on the quality of the memory device and the data
transmission.
Data is stored partly or not at all if for example:
The power supply of the external memory device is interrupted during the write process, or
The external memory device is prematurely disconnected from the photometer during the data backup.
To prevent a data loss we recommend the following:
Save all data internally in the photometer first.
After performing a backup leave the USB memory device connected to the photometer for some time.
Check whether the stored data is complete, e.g. on a PC.
Use the USB memory device for data transport but not for permanent data storage.
1.55

MEASURMENT DATASETS

1.55.1
E

LEMENT OF A MEASUREMENT DATASET

A complete measurement dataset consists of:
- Consecutive number (is automatically assigned by the spectrophotometer).
- Date/time.
- Identification (e.g. ID or "AutoStore").
- User name.
- Measured parameter, e.g. method number, dilution, wavelength (depending on the measuring mode).
- Measured value with unit and if necessary citation form.
1.55.2
O
PÉRATIONS WITH MEASUREMENT DATASETS
Measurement datasets can be
Stored
Displayed and printed
Filtered, i.e. selected or hidden based on certain criteria
Deleted
Copied
Saving and exporting
Measurement datasets of these modes are saved with <STORE> or
AutoStore in the measured value memory of the spectrophotometer first
(5000 locations).
The measured value memory is available via the Measurement data
memory menu.
Here it is possible to view or filter stored measurement datasets or
export them into a PC readable file (*.csv) (<STORE>).
Csv files of these modes cannot be imported back into the photometer.
Measurement datasets of these modes are saved and exported as a PC
readable file (*.csv) with <STORE>.
Csv-files of these modes can be re-imported and viewed on the
photometer.
- 76 -
UviLine Series

Publicité

Chapitres

Table des Matières
loading

Table des Matières