UviLine Series
It is the number of figures after comma for the result.
* Unit
i.e.: mg/L (10 characters maximum).
* Citation form
(18 characters maximum).
Enter the chemical formula of the compound to be measured. i.e.: NO
3
* Designation
(18 characters maximum).
Enter the compound name. i.e.: Nitrate
* Version
(18 characters maximum).
Enter the parameter concentration range. i.e.: 0.9 – 75.3 mg/L
* Calibration curve
There are several possibilities:
- Measure the standard solutions.
- Enter the equation of the calibration curve or the factor value.
- Simulate measurements while entering of the couples of Absorbance/Concentration values
* Cell changer mode
Usefull if the instrument is equiped with a cell changer
* Standard ID
Enter the identification number of the standards if this one exists.
* Standard manufacturer
Enter the standard manufacturer name if this one exists
* Blank value
YES or NO.: Measure done or not of the reagent blank
* Standard conc 1
Enter the concentration of standard 1.
- 36 -