UviLine Series
- It is also possible, if the solution is unstable, to measure it again several times. The software will make the average of the
various measured values.
Remark: The number of times to measure again the solution is not limited.
- Press, then F4 key [Apply].
- The operator can choose to preserve the last measured value by validating the option "Last measured absorbance" or to
use the average value calculated by choosing and validating the option "Median".
- Install the first standard to be measure. Then press F3 key [Measurement].
- Start measurement of the first standard by pressing <START ENTER> key.
- As for the blank, the operator can reject measurement (F2 key [Discard]) or make several times the measurement of the
same standard (F1 key [Next meas.]) and preserve the average value.
- Make, in the same way, the measurement of the other standards.
-Press F4 key [Next.] to proceed
NOTE:
If the programming of standards concentration is not increasing or decreasing in a monotonous way or if the measurement
of the blank and standards absorbances is not, also increasing or decreasing in a monotonous way (see example below),
the spectrophotometer will display an error prompt (see below)