Function
Application
4 – 20 mA
Analogue output 4 – 20 mA
Relay test
Relay function tests
Analogue
Analogue output tests
test
11.3.3 Settings
For the outputs A1 ... A12, one of the above-listed functions can be set in the submenu
Operating setting/Output. Outputs for which no output is available because no output module is
fitted are available for internal control tasks as so-called virtual outputs, e.g. for the control of tariff
registers.
The following table shows settings in / information on the submenu Operating setting/Output
It illustrates two cases: one with an assembled output module and the other with virtual output (i.e.
non-assembled output module):
Case 1
with output module
Output
1
Mod-No 3
Trm.no. 86 – 50 - 51
Funct.
Status
...
Lot no. 96370508
HW no. 1804169
FWV
V 01.00.03
Case 2 applies if the virtual output for internal control functions is used, e.g. to control a tariff meter.
11.3.4 Pulse function
Using the pulse function allows transmission of the meter increments as pulse sequence. The follow-
ing values are displayed in the menu:
Output 1
...
Funct.
Impulse
Outvar Energy
P.val
1.000 kWh
t puls
10 ms
Pol.
Contact
...
24
Case 2
Virtual output
Output 2
Mod-No no module
Trm.no. no module
Funct.
Limit value
...
Batch No. 0
HW_Nr. 0
FWV
no module
Output variable of
calc. 1 is energy
Pulse value
Pulse width
Switching behav-
iour
Note
Analogue output, interrupt can be detected by
module with analogue input
Switches the relay to the displayed status.
Setting of a target value and measuring of the
actual value with an ammeter.
Explanation
Select the output with the
Display of the module no.
Display of the terminal numbers
Output function, see table above
Function-dependant parameters, see description
under the respective function
Production batch number
Hardware version number
Firmware version number
Select the output with the
Pulse output function
Input via double list. Only activated calculators are
displayed. The available meter values depend on the
calculator function (e.g. E, V, M)
Input range: 0.00001 to 999999
See below for specification of the unit
10 ms, 50 ms, 500 ms or 1000 ms
Duty cycle: 50%
Make contact: A pulse closes contact
Break contact: A pulse opens contact
CALEC
keys
keys
®
energy master, Output 24V