P9.6 Select Parameters
Function available only for PC 60 VioLab.
Through this setup menu, it is possible to select which parameters do NOT display in measure mode.
Access menu P9.6. The icon
YES: in measure mode the pH parameter is kept active
NO: in measure mode the pH parameter is not displayed
Confirm the selection with key
mV parameter and then for all the parameters up to TDS
Example: The user wishes to work with pH, Conductivity and TDS parameters only.
In the P9.6 menu: pH -> YES / mV -> NO / ORP -> NO / COND -> YES / TDS -> YES
Press twice key
to return to measure mode. Scrolling with the key
and TDS are shown.
Note:
At least one of all the parameters must be enabled with YES
P9.8 Reset Settings
Access this setup menu to restore the instrument to factory conditions.
IMPORTANT: Restoring the factory parameters does not delete the stored data.
DataLink+ Software
It is possible to connect the instruments of the 70 Vio series to the PC and then use the DataLink + 1.6
software (and later versions) to perform data download, Data Logger directly on PC and exports in .xls (Excel)
and .pdf. The software can be downloaded for free from the web site (pay attention to the correct
installation of the drivers).
https://www.giorgiobormac.com/it/download-software_Download.htm.
Connect the USB cable inside each package to the USB port on the top panel of the instrument and the
other end to a COM port on the computer.
Use the USB cable supplied with the instrument only.
Start the program and then switch on the instrument.
Wait for connection to be established (the connection data are shown at the bottom left of the display).
 Functions
Download: the data saved in the instrumental memory are downloaded to a PC and displayed in the
table for processing.
M+: instantaneous acquisition of a value (equivalent to the manual Data Logger option).
Logger: automatic acquisition with set frequency.
Empty: emptying the data in the table. If the password is active, it will be requested.
Export to Excel / Export to PDF: export to PDF and Excel of all the data in the table, of graphs,
calibration reports and instrumental information.
Save to file / Open from file: saving the data in the table and possibility to reload them in order to
process them or continue recording.
Select the language: set the interface language (Eng – Ita – Deu – Esp – Fra – Cze).
Table / Graph: how to display the acquired data. The graphs are divided by parameter and can be
printed separately.
UM Serie 60VioLab rev.0 20.05.2020
flashes, with directional keys choose:
; now the icon
flashes, then repeat the same operation for the
(per Windows 7/8/XP/10)
33
.
only the parameters pH, COND
EN

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Pc 60 violabCond 60 violab60 violab serie

Table des Matières