Calibration With Manual Values - XS Instruments pH 60 VioLab Manuel D'utilisation

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

Press the key
The measured value flashes on the display and then the icon of the pH 7.00
beaker appears at the bottom left
calibrated on the neutral point.
Remove the electrode, rinse with distilled water and dab gently with paper
towel. Dip the sensor in the pH 4.01 buffer solution ("CHANGE BUFFER").
The instrument is now ready to recognize the second calibration point.
Next to the string "2ND POINT PH", the different buffers, that the device can recognize automatically,
scroll.
When the 4.01 value is recognized and the icon
string "PRESS OK".
The actual measured value and the Slope% flash on the display; subsequently, the icon of the beaker
pH 4.01
appears next to the green beaker, indicating that the instrument is calibrated in the acid
field.
Remove the electrode, rinse with distilled water and dab gently with paper towel. Dip the sensor in the
pH 10.01 buffer solution ("CHANGE BUFFER").
The instrument is now ready to recognize the third calibration point.
Next to the string "3RD POINT PH", the different buffers, that the device can recognize automatically,
scroll.
When the value 10.01 is recognized and the icon
the string "PRESS OK".
Switching from an acidic to a basic pH may take a few more seconds to achieve stability.
The actual measured value and the second Slope% flash on the display; subsequently, the icon of the
beaker pH 10.01
calibrated in the alkaline field.
At the end of the third calibration point, the instrument returns automatically to measure mode.
To perform a one- or two-point calibration, press the key
Note: electrode calibration is an essential operation for the quality and truthfulness of a
measurement. Therefore, make sure that the buffers used are new, unpolluted and at the same
temperature.
ATTENTION: Before proceeding with the calibration operations, carefully consult the safety data sheets
of the substances involved:
Calibration buffer solutions.
STORAGE solution for pH electrodes.
Filling solution for pH electrodes.
 Calibration with manual values
Example
two-point calibration pH 6.79 e pH 4.65 (DIN19267)
:
Access the Setup menu for pH and select in P1.1 Custom, press twice the key
measurement and position in pH mode
Press
to enter the calibration mode.
Rinse the electrode with distilled water and gently dab with paper towel. Dip the electrode in the first
pH buffer solution (e.g. pH 6.79).
UM Serie 60VioLab rev.0 20.05.2020
as indicated by the string "PRESS OK".
indicating that the instrument is
appears next to the green and red beakers, indicating that the instrument is
.
18
appears, press the key
appears, press the key
once finished the first or second point.
as indicated by the
, as indicated by
to return to the
EN