Plusmed pM-7E A Mode D'emploi page 14

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 22
Note: Apply small amount of distilled water around the part (pressed into the holder mouth) of
holder plug during installing, which is good for tightly pressing the holder plug and enhancing its
sealing.
Tube Connecting Diagram
1.Suction tube 2.Air filter 3."IN" mark 4.Suction conductor 5. Liquid holder 6.To the phlegm
suction catheter
3. Power line connection
Connect the plug with the power source. Turn on the power supply, and the power indicator will
illuminate.
Note: The power plug is used for power shut-off, and the power socket shall be grounded reliably.
4. Connector inspection
• Turn tightly the negative pressure regulating valve clockwise, and block the air suction inlet with
the finger or the rubber head of dropper, or fold up and hold the suction tube;
• Start the aspirator for running with no strange sound; the pointer on the vacuum meter will
quickly reach up to the limit negative pressure. Release the suction inlet, the pointer will return
below 150 mmHg. If so, the connector can be regarded as being in good connection.
• Attach the phlegm suction catheter. The negative pressure in the negative pressure system shall
be less than 450 mmHg at the time of attaching F6 suction catheter, less than 300 mmHg when
attaching F8 suction catheter and less than 225 mmHg when attaching F12 suction catheter. If
so, the phlegm aspirator is considered as being in normal condition.
14
1
3
2
1
4
6
5