Working temperature
(Environment of measuring device) . . :0°C to +50°C (32°F to 122°F)
Storage temperature. . . . . . . . . . . . . . :-10°C to +60°C (14°F to 140°F),
battery removed)
Relative humidity . . . . . . . . . . . . . . . . :0 to 80 %, not condensing
Temperature for
guaranteed accuracy . . . . . . . . . . . . . : +23°C ±5 K
Temperature coefficient . . . . . . . . . . . : in addition 0.01% of reading +0.03°C
Battery replacement indicator . . . . . . : „
Battery type . . . . . . . . . . . . . . . . . . . . :NEDA 1604 9V or 6F22 9V (alkaline)
Weight. . . . . . . . . . . . . . . . . . . . . . . . . :210 g (with battery)
Dimensions (L x W x H). . . . . . . . . . . . :184 x 64 x 30 mm (without cables)
Measuring tolerances
Accuracy specified in + (% of reading + indication error in Kelvin "K") "K" for Kelvin is
used as the absolute value of a differential temperature or temperature deviation.
Accuracy for 1 year at a temperature of +23°C ±5K, at a relative humidity of less than
80%, non-condensing. The warm-up time is 1 minute
Measuring range
Measuring device:
-200°C bis +200°C
+200°C bis +400°C
+400°C bis +1370°C
-328°F bis -200°F
-200°F bis +200°F
+200°F bis +2498°F
Temperature sensor TP-K01
-50°C to +200°C
-58°F to +392°F
If the max. admissible input quantities are exceeded, this might
under adverse circumstances damage the measuring device, or
endanger the user's life.
(or 0.01% of reading +0.06°F) per K in the
range from 0°C to 18°C
and from 28°C to 50°C
„ at less than approx. 7.3 V
+
–
battery voltage
Accuracy
±(0,2%+1°C)
±(0,5%+1°C)
±(0,2%+1°C)
±(0,5%+2°F)
±(0,5%+2°F)
±(0,5%+2°F)
±2.2K or ±0.75%
±3.6K or ±0.75%
Resolution
0,1°C
1°C
1°C
0,1°F
0,1°F
1°F