WIKA IL-10 Mode D'emploi page 11

Transmetteur de pression immergeable à sécurité intrinsèque
Masquer les pouces Voir aussi pour IL-10:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

3. Safety
Model
Model IL-10, with FEP cable
1) The respective ambient and medium temperatures are limited by:
The maximum permissible surface temperature, valid for applications that require EPL Ma (150 °C)
Temperature class assignment, valid for gas applications that require EPL Ga or Gb (maximum ambient temperature)
The permissible power, Pi, valid for dust applications that require EPL Da or Db (maximum ambient temperature)
Cable properties (minimum and maximum ambient temperature)
WIKA operating instructions submersible pressure transmitter model IL-10
EPL
Ambient and medium
temperature
Ma
-30 ... +105 °C
Ga, Ga/Gb, Gb
-30 ... +60 °C
-30 ... +80 °C
-30 ... +105 °C
Da, Da/Db, Db
-30 ... +40 °C (750 mW)
-30 ... +70 °C (650 mW)
-30 ... +100 °C (550 mW)
Temperature class, surface
temperature
1)
not applicable
T6
T5
T4
120 °C
120 °C
120 °C
EN
11