WARNING, risk of DANGER! The operator agrees to refer to these
instructions whenever this danger symbol appears.
Equipment protected throughout by double or reinforced insulation.
Must not be applied to or removed from conductors at dangerous
voltages. Type B current sensor as per IEC 61010‑2‑032.
The CE marking guarantees conformity with European directives and
with EMC regulations.
The rubbish bin with a line through it means that in the European
Union, the product must undergo selective disposal for the recycling
of electric and electronic material, in compliance with Directive WEEE
2002/96/EC.
Battery.
Definition of measurement categories:
Measurement category IV corresponds to measurements taken at the source of
low‑voltage installations.
Measurement category III corresponds to measurements on building installa‑
tions.
Measurement category II corresponds to measurements taken on circuits directly
connected to low‑voltage installations.
Measurement category I corresponds to measurements taken on circuits not
directly connected to the network.
Thank you for purchasing a MiniFLEX MA100 flexible current sensor. To obtain
the best service from your unit:
read these operating instructions carefully,
comply with the precautions for use.
This instrument is protected against voltages of not more than:
600V with respect to ground in measurement category III or 300V in CAT IV on
the terminal or output cord side.
1000V with respect to ground in measurement category III or 600V in CAT IV
between the sensor and the conductor of which it measures the current.
The peak maximum rated current is equal to 1.5 times the upper range limit.
The protection provided by the instrument may be impaired if the instrument is used
other than as specified by the manufacturer.
ENGLISH
PRECAUTIONS FOR USE
18