Appropriate Usage

The accessories may only be used under those
conditions and for those purposes for which it
was conceived. For this reason, safety refer-
ences, technical data (including environmental
conditions), and the usage in dry environments
must be followed.
If probes are modified by the user, operational
safety is no longer ensured.
Accessories-handle surface, refer to
figure 2 (page 2)
Only touch test leads and test probes at handle
surface provided. Never directly touch test probe
tips or touch test probes beyond the barrier.

Fuse Replacement

The fuse can be checked with a simple continuity
test.
Prior to fuse replacement, disconnect the acces-
sory (cable or probe) on both side every time.
Exclusively use fuses of voltage and current
values in compliance with the technical data
section. Using auxiliary fuses, in particular short-
circuiting fuse holders is prohibited and can
cause probe destruction or serious bodily injury
of operator.
Completely unsrew the fuse holder (FH). See figure 3
(page 2)
Remove defect fuse.
Push the fuse into the fuse holder.
Screw in fully the fuse holder to the end stop.