7.5.2. Operational test

x
This test must be performed prior to each sterilisation.
Connect the contra-angle to the motor ; point the head of the contra-angle downwards.
Operate the contra-angle slowly for 30 seconds in order to get rid of the excess oil,
then at full speed for 30 seconds. Wipe the contra-angle with a disinfectant wipe if the
oil has dripped.
NOTE : if you notice a rise in temperature, irregularities, vibrations, or abnormal
noises during operation of the contra-angle, contact your approved technical service
immediately.
7.6. Sterilisation
• We strongly recommend the use of an autoclave with pre-vaccum cycle.
• Any other method of sterilisation should be avoided.
• Read the instruction leaflet provided by the steriliser manufacturer.
• Remove the rotating instrument from the contra-angle prior to sterilisation. (See §6.3)
• Only sterilise instruments which have previously been disinfected, cleaned, lubricated
and tested.
• Use sterilisation pouches suitable for the contra-angle and the steriliser, in
accordance with the standard NF EN 868. Always use one single contra-angle per
pouch.
• Adhere to the space specified between pouches and do not overload the steriliser.
• Make sure that the device does not have any areas of corrosion or cracks, and check
that it is operating properly. Ensure that the product is dry ; if necessary, dry any
residual water with medical quality pressurised air.
• After each sterilisation cycle, check that there is no water remaining on the inside
and outside of the packaging. Make sure that the flow indicator has changed to the
correct color.
• In order to avoid any retention of water, place the pouch in the steriliser in such a way
that any concave parts are face down.
• Keep the devices in the sterilisation pouches away from light, moisture and any
contamination. Follow the manufacturer's recommendations as seen on the packaging.
• The duration for which the device is kept after sterilisation should not exceed 1
month. Label the devices, specifying the expiration date.
After the expiration date, repeat the cleaning and sterilisation procedure.
g
• The contra-angles should be sterilised
before they are first used and following
each use.
• The device should be sterilised at 135°C at 2.13
bars (275°F at 30.88 psi), in a steam steriliser only,
for a minimum duration of 20 minutes (time for
which sterilisation must be continued).