Error Codes (Rt 5/10/15); Accessories - IKA RO 5/10/15 Mode D'emploi

Table des Matières

Publicité

Error Codes (RT 5/10/15)

Error- Cause
code
Er 3
Temperature inside device too high
Er 6
Break in safety circuit
Er21
Safety relay doesn't open during test
Er22
Safety temperature is too low during test
Er25
Heating and switching element monitoring
Er26
Difference between temperature of safety
sensor and temperature of control sensor
control temperature > (safety tempe-
rature + 40 K)
Er31
Triac error detection doesn't work
Er46
Difference between temperature of safety
sensor and temperature of control sensor
control temperature > (control tempe-
rature + 40 K)
If the actions described fail to resolve the fault or another error code is displayed then take one of the following steps:
- Please contact the service department;
- Send the device for repair, including a short description of the fault.

Accessories

• Stirring bar:
ø 8mm; lenght 30 mm
• RSE
Stirring bar remover
Effect
Solution
Heating off - Switch off device and allow to cool down
Heating off - Plug in contact plug (L)
- Plug in contact thermometer or PT 1000 temperature sensor
- Replace faulty connection cable, plug, or contact thermometer
- Switch off device
- Warning! Only to be carried out by authorised service personnel:
Check PCB
Heating off - Switch off device
- Warning! Only to be carried out by authorised service personnel:
Check PCB and safety temperature sensor
Heating off - Check the external temperature controller (f.e. Er 5 ETS-D)
- Switch off device
- Warning! Only to be carried out by authorised service personnel:
Carry out an internal test on the device to check the plug-in connector
for the heating element, the safety temperature sensor or the PCB
Heating off - Switch off device
- Warning! Only to be carried out by authorised service personnel:
Carry out an internal test on the device to check the plug-in connector
for the temperature sensor
Heating off - Switch off device
- Warning! Only to be carried out by authorised service personnel:
Check PCB
Heating off - Switch off device
- Warning! Only to be carried out by authorised service personnel:
Carry out an internal test on the device to check the plug-in connector
for the temperature sensor