5-9 Handpiece Stand
When the Handpiece is not in use, place the Handpiece horizontally on the Handpiece stand.
Make sure that the Tip does not rest on the Handpiece Stand.
Make sure that the Handpiece rests horizontally on the Handpiece stand, and not at an angle.
CAUTION
To maintain an hygienic condition, the Handpiece Stand should frequently be cleaned with alcohol.
No Tip contact
6
Check before treatment
Before use, operate the product without patient contact and check the following points. If any abnormality, such as low
vibration, noise, heat, is detected at anytime disconnect the product and contact your Authorized NSK Dealer.
-Is irrigation solution flowing correctly at the Tip?
-Is the Tip vibrating correctly?
-Are there any abnormalities at the Tip, Eg. vibration, noise and heat?
-Is the Handpiece illumination operating correctly?
7
Operating method
7-1 Check Irrigation
1) Confirm that the Irrigation Tube correctly connected to the Coolant Solution Pack, the Irrigation Pump and the
Handpiece.
2) Turn ON (-side) the Power Switch.
3) Open the Tube Clamp which is placed between the Irrigation Needle and the Irrigation Pump. Then, depress the
Ultrasonic ON/OFF Button on the Foot Control and confirm that the irrigation solution correctly flows from the Tip.
When a new Irrigation Tube is used, it may take several seconds before irrigation solution flow commences.
7-2 Set the Power Level
1) Change the Mode – Use MODE key.
2) Select the Program No. – Use PROGRAM key.
3) Adjust the Power Level – Use POWER key.
4) Adjust the Coolant Flow Level – Use COOLANT key.
5) Adjust the Handpiece illumination intensity
– Use LIGHT key.
6) Select the Burst Output Level – Use BURST
key. (In SURG Mode only).
* Power outputs at various Modes are
compared in Fig. 20.
Horizontal
SURG
10%
ENDO
5%
100%
00%
PERIO
5%
100%
13
Not horizontal
Fig. 19
150%
OUTPUT
Fig. 20