BODY MOTION SYMBOL
Blood pressure value taken while moving cannot be said to be the correct value
because body movement can affect blood pressure.
This product analyzes pulse wave and displays [
detected. [
] indicates the results might be affected by body movement.
IRREGULAR PULSE RHYTHM SYMBOL
IRREGULAR
PULSE
RHYTHM
INDICATION shows that the pulse
waves are not appropriately detected
during blood pressure measurement.
When [
] is indicated, please
take a measurement again in a
calm condition as the blood
pressure measurement result may
not be accurate.
* Irregular pulse rhythm can be caused by talking, moving or the health condition etc.
MEMORY FUNCTION
The measured values are automatically saved for later review in either one of
two memory banks, [1] or [2]. These two memory banks can be used to save
readings of two persons separately or to save morning readings and evening
readings separately.
Each bank can save up to 60 readings. When the number of stored readings
reaches 60, the oldest reading will be deleted to record new reading.
Saved readings are recalled with measurement date and time when clock
function is activated. *The clock needs to be activated before measurement for
date and time to be saved with the measured valued.
[Err] results are not saved.
] when body motion is
regular pulse rhythm
An example of irregular pulse rhythm
compared to regular pulse rhythm
CARE AND MAINTENANCE
Keep the product clean. Inspect its cleanliness after use. To clean, use only
a soft dry cloth. Do not use gasoline, paint thinner, or other strong solvents.
Since the cuff may absorb perspiration and other fluids, inspect it for stain and
discoloration after each use. When cleaning the cuff, use a synthetic detergent
and gently rub the surface.
When storing the product, do not place heavy objects on it. Do not fold or bend
the cuff forcibly. In case the product is stored in the environment with ambient
temperature above 40˚C or below 10˚C, please leave it for at least 2 hours
before taking a measurement. Remove the batteries if the product is to be stored
for an extended period of time. Keep the batteries out of reach of children.
Stop using the product immediately and contact your dealer or the manufacturer
in case any visible damage is found on the device.
We suggest that you have your blood pressure monitor checked every 2
years. This operation may only be performed by the manufacturer or by firms
authorized by the manufacturer.

PRODUCT SPECIFICATIONS

Model
: WSK-1011
Operating Principle
: Oscillometric method
Indicator
: 15 digits liquid crystal display
Pressure indicating range : 0 to 300 mmHg (cuff pressure)
Measuring range
: 50 to 250 mmHg (systolic), 40 to 180 mmHg
(diastolic), 40 to 160 bpm (pulse rate)
Accuracy*
: ±3 mmHg (cuff pressure), ±5% of reading (pulse rate)
Inflation
: Automatic inflation (FUZZY logic)
Deflation
: Automatic (electric control valve)
Exhaust
: Automatic exhaust valve
Power Supply
: Two 1.5 volt AAA alkaline (LR03) type batteries
-4-