TECHNI CAL DE S CR I PT I O N F O R EL ECT R OM A G N E T IC D IS T U R B A N C E S
WSK-1011 complies with the Electromagnetic Disturbances standard,
IEC60601-1-2:2014.
As a medical electrical equipment, special precautions regarding the
electromagnetic disturbances shall be taken at usage of the device according to
the information provided below.
• The device is not intended for use in environments where the intensity
of electromagnetic disturbance is high, such as near active HF surgical
equipment and MRI (magnetic resonance imaging) equipment etc.
• Use of the device adjacent to or stacked with other equipment must be
avoided because it could result in improper operation.
• Use of accessories other than those specified or provided by the
manufacturer could result in increased electromagnetic emissions or
decreased electromagnetic immunity of the device and result in improper
operation.
• Portable RF communications equipment (including peripherals such as
antenna cables and external antennas) should be used at least 30cm away
from any part of the device. Otherwise, degradation of the performance of
this equipment could result.
Please contact your dealer or the manufacturer for specific information regarding
the compliance to the standard.
-15-