is displayed.
The cuff is not properly applied. Properly apply the cuff. If Err-2
appears after re-trying to apply the cuff properly, assume the cuff is
damaged. Contact your dealer.
is displayed.
Batteries are weak. Replace all batteries with new ones.
is displayed.
START/STOP KEY was being touched at installation of batteries. Turn
off the monitor once with "START/STOP" KEY.
Nothing is displayed.
???
Batteries are depleted. Replace all batteries with new ones.
Batteries are inserted incorrectly. Reinsert the batteries correctly.
Battery terminals are not clean. Clean the terminals with dry cloth.
AC adaptor is not connected. Connect AC adaptor.
Touching the keys with tip of finger or nail is touching the key. Touch
the key with the plane of your finger.
Clock is not displayed.
???
The clock is inactivated. NOTE: Clock is inactivated when batteries are
removed. Adjust date and time and activate the clock.
Measurement date and time are displayed with [
???
The clock is not activated. Adjust date and time and activate the
clock.
Measurement was taken before the clock was activated. Measurement
date and time cannot be saved without clock being activated.
If you cannot get correct measurement with the methods above, contact your
dealer. Do not disassemble or tamper with the internal mechanism.