Front Panel - Apex DOMUS 4 Instructions D'utilisation

Masquer les pouces Voir aussi pour DOMUS 4:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 60
English
2.2 Pump Unit
1
1
2

2.3 Front panel

1. Pressure Adjust Levels
Pressure adjust levels controls the air pressure output. When turning up (
pressure will increase. Vice versa (
will support the heavier weight patient. In order to find the correct pressure setting: First inflate
the mattress to max pressure level, have the patient lie on top of the mattress, then decrease
the pressure gradually until the patient feel most comfortable without bottoming out. Have the
caregiver insert palm facing up and place between patient's buttock's area and static cell.
Should the palm able to touch patient's buttock, increase the pressure until the skin is least
1/2" inch away from the palm. Always leave at least 1" inch space between patient's buttock
areas to the static cell under to prevent bottoming out.
2
Front
1. Quick Connectors
2. Front Panel
3
Rear
1. Power Switch
2. Power Cord
3. Mounting Brackets
2
3
) for decreasing air pressure. Higher-pressure output
5
1
), the output