4. Operations
NOTE: Always read the operating instructions before use.

4.1 General operation

4.2 Emergency CPR Operations
CPR must be performed on a firm surface. Therefore, if an
emergency CPR situation occurs with the patient on the
mattress, the mattress must be quickly deflated. This is
done by quickly pulling on the CPR tag located at the head
of the mattress on the patient's right-side. The quick
connector found from the pump unit can also be
1. Switch on the main power switch on the side of
the pump.
2. The pump unit is to deliver air into mattress,
wait at least 30 minutes for mattress to be
inflated.
3. The low-pressure indicator (yellow LED) will
light up when the mattress is not fully inflated.
4. When the appropriate pressure is reached, the
low-pressure indicator (yellow LED) will go off
and pump will enter the alternating mode.
5. According to the weight and height of the
patient, adjust the pressure setting to the most
comfortable level without bottoming out, then
the pressure in mattress will slowly increase to
the intended value after the air mattress is
ready to use.
NOTE: Every time when mattress is first
setup for use, It will be forced to
execute Maxfirm for the quickest
inflation. User can then adjust air
mattress to the desired softness
afterward.
9