8. Replacing the O-rings
Replace the O-rings if water is present in the exhaust air line. This is an indication of possible water leakage
within the motor. ALWAYS change the complete set of O-rings.
1) Gently remove each O-ring by hand.
2) Insert the complete set of new O-rings in the correct grooves.
* Refer to Spare Parts List to identify the correct parts.
CAUTION
· Do not force the new replacement O-ring with excessive pressure.
· The blue O-Ring is thinner than the 3 black O-Rings. When inserting new O-rings, make sure they are inserted in the correct grooves as shown
in the Fig. 8.
10
O-ring (Blue)
9. Periodical Maintenance Checks
Motor Insert
Perform periodical maintenance checks every three months, referring to the check sheet below. If any abnormalities are found, contact your Authorized
NSK Dealer.
Points to check
O-ring (Black)
Rotation
Fig. 8
10. Symbol
This product is Autoclavable up to Max.135°C.
Conforms to CE European Directive of "Medical equipment directive 93/42/EEC."
Manufacturer.
Caution: U.S. Federal law restricts this device to sale by or on the order of a licensed physician.
Details
Rotate the motor and check for abnormalities such as abnormal rotation, vibration, noise, and overheating.
This product can be washed via Thermo Disinfector.
Authorized representative in the European community.