adjuST CaLIBRaTION – advaNCEd OPTION
IMPORTaNT: Always obtain three (3) test results from the grain elevator
for the sample being compared. Average these three (3) results. Compare
this average with the average of three (3) tests by the moisture tester.
1. Each grain scale can be individually
adjusted up to 5.0% by increments
of 0.1% to more closely match the
results of an elevator tester.
2. First select the grain to be adjusted.
3. Test the grain selected, using the
sample you wish to adjust to. Once
valid moisture appears, press the CAL
button (A).
NOTE: Calibration of given grain
cannot be performed unless a valid
moisture test has been performed.
4. Once CAL button (A) is pressed,
tester will now display the obtained
moisture reading and the current
offset applied to that moisture range.
5. Press the UP arrow (B) to raise the adjustment amount or press the
DOWN arrow (C) to lower. The tester will add or subtract up to 5.0%
by increments of 0.1% to the current moisture range.
6. After the adjustment amount has
been selected, press the CAL button
(A) to return to grain tester mode.
NOTE: Adjustment can NOT be
made to factory calibration to
produce moisture readings in a
tester that shows "BELOW LIMIT" or
"ABOVE LIMIT" readings.
IMPORTaNT: This tester incorporates
Multi-Point Calibration for each grain.
Therefore, once a valid test is taken and
an adjustment is made, the adjustment
will only affect the moisture range of the
sample tested.