adjuST CaLIBRaTION – SIMPLE OPTION
IMPORTaNT: Always obtain three (3) test results from the grain elevator
for the sample being compared. Average these three (3) results.
Compare this average with the average of three (3) tests by the moisture
tester before making any adjustments.
1. Each grain scale can be individually
adjusted up to 5.0% by increments
of 0.1% to more closely match the
results of an elevator tester.
2. First select the grain to be adjusted.
3. Press CAL Button (A), tester will now
display the current offset applied to
the grain.
4. Press the Up Arrow (B) to raise the
adjustment amount or press the Down
Arrow (C) to lower. The tester will add
or subtract up to 5.0% by increments
of 0.1% to the calibration offset.
5. After the adjustment amount has been
selected, press the CAL Button (A) to
return to grain tester mode.
NOTE: Adjustment can NOT be made to factory calibration to
produce moisture readings in a tester that shows "BELOW LIMIT" or
"ABOVE LIMIT" readings.
IMPORTaNT: When using the SIMPLE
calibration offset option, the tester will
apply calibration offset to selected grain
regardless of moisture range.
Operation
05-10