Instruction Manual; Intended Use - Mettler Toledo Optical O2 Simulator Manuel D'instruction

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 38
en
20
1

Intended Use

With the Optical O
simulate the output of different digital optical
oxygen sensors. The simulator is connected
to the transmitter instead of the real sensor to
verify connections are correct, and to simulate
the behavior of the transmitter in different mea-
surement conditions and in case of a suspected
sensors error.
Depending on the chosen condition mode, the
simulator works as if it is measuring in air or in
an oxygen-free medium.
As soon as the simulator is connected to the
transmitter, the transmitter will show pre-de-
fined data such as DLI, ACT, operating hours,
autoclaving cycles, SIP and CIP of the selected
sensor.
The Optical O
lowing tasks:
• verifying an optical O
sor, cable and transmitter)
• checking transmitter settings are correct
• troubleshooting.
The simulator is designed for use in the process
industries and can be used in combination with
M400 and M800 transmitter.
Optical O
Simulator
2
30 422 249
en

Instruction manual

Simulator is intended for the fol-
2
Simulator it is possible to
2
measuring loop (sen-
2
© 02 / 1 8 Mettler-Toledo GmbH
Optical O
Simulator
2
Printed in Switzerland

Publicité

Table des Matières
loading

Table des Matières