The battery life depends on the ranges used
10 A range
1 A range
100 mA range
10 mA range
Instrument on standby or off
(1) established for measurements lasting 5s, every 25s.
6.7. ENvIRONMENTAL CONDITIONS
Use indoors or outdoors.
Range of use
Storage (without battery)
Altitude
Degree of pollution
For long-term storage (2 years) with the battery, conditions must not depart from the range -20 to
+30°C and 85% RH; otherwise, the battery life will be degraded. For short-term storage (1 month), the
temperature can reach 50°C.
6.8. CHARACTERISTICS OF CONSTRUCTION
Overall dimensions of the instruments (L x W x H): 273 x 247 x 176 mm
Mass: approximately 4.5 kg
IP 53 per NF EN 60529 (Ed. 92)
IK 04 per NF EN 50102 (Ed. 95)
6.9. CONFORMITY TO INTERNATIONAL STANDARDS
Electrical safety as per EN 61010-1 (Ed. 2 of 2001)
Measurement according to EN 61557 (Ed. 2 of 2007) parts 1 and 4.
Safety level categories: measurement category III, 50V with respect to earth, 500V differential between
terminals, and 300V cat II on the charger input
6.10. ELECTROMAGNETIC COMPATIBILITY
The instrument satisfies the CEM and DBT directives required for the CE marking and product standard
EN 61326-1 (Ed. 97) + A1 (Ed. 98)
Emissions in residential environment
Immunity in industrial environment
Number of measurements (1)
850
3 500
4 500
5 000
battery life 4 to 6 months
- 10 to +55 °C
10 to 85 % RH
- 40 to +70 °C
10 to 90 % RH
< 2000 m
2
0