Environmental Conditions; Characteristics Of Construction - Chauvin Arnoux C.A 742 IP2X Notice De Fonctionnement

Masquer les pouces Voir aussi pour C.A 742 IP2X:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 1

4.3. ENVIRONMENTAL CONDITIONS

The device is of type N. It must be used under the following
conditions:
%RH
100
90
80
70
60
50
40
30
20
10
0
-50 -40 -30 -20 -10 0
1: Reference domain
2: Operating range
-15 to +45°C and 20 to 95% RH without condensation.
(35°C max. at 95% RH)
3: Range in storage (without batteries)
-40 to +70°C and 20 to 95% RH without condensation.
If an extended period of non-use is anticipated, or for storage,
withdraw the batteries from the housing.
For use indoors and outdoors.
Pollution degree: 2.
Altitude: <2000m.
4.4.
4.5. ALIMENTATION
The C.A 742 IP2X is powered by two 1.5V alkaline batteries
(type AAA or LR3).
The battery life is 7,000 10-second measurements.
The batteries can be replaced by rechargeable batteries, but
the life between charges will be much shorter.

4.6. CHARACTERISTICS OF CONSTRUCTION

Dimensions (LxWxH)
Weight
Cord
Protection class
Drop
3
2
163 x 64 x 40mm
approximately 210g
length 90cm
IP65 per IEC 60529
IK04 per IEC 50102
2 metres
1
10 20 30 40 50 60 70 80 90
34
°C

Publicité

Table des Matières
loading

Table des Matières