Environmental Conditions - Chauvin Arnoux C.A 762 Notice De Fonctionnement

Masquer les pouces Voir aussi pour C.A 762:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 1
3.2.3. PHASE IDENTIFICATION
15Hz<frequency<65Hz
50 V
<voltage<690 V
ac
150 V
<voltage<690 V
ac
3.2.4. PHASE-ORDER
Frequency between 45 and 400Hz.
Voltage between 50 and 690 Vac between phases.
Time to acquisition of the information after contact ≤1s.
Retention time of the information: 10s.
Acceptable amplitude unbalance: 20%.
Acceptable level of harmonics in voltage: 10%.
Rejection of power company remote control signals (TCC-
175Hz-188Hz).

3.3. ENVIRONMENTAL CONDITIONS

The device is of type N. It must be used under the following
conditions:
%RH
100
90
80
70
60
50
40
30
20
10
0
-50 -40 -30 -20 -10 0
1: Reference domain
2: Operating range
-15 to +45°C and 20 to 95% RH without condensation.
(35°C max. at 95% RH)
3: Range in storage (without batteries)
-40 to +70°C and 20 to 95% RH without condensation.
If an extended period of non-use is anticipated, or for storage,
withdraw the batteries from the housing.
For use indoors and outdoors without rain.
Pollution degree: 2.
Altitude: <2000m.
3.4. ALIMENTATION
The C.A 762 is powered by two 1.5V alkaline batteries (type
AAA or LR3).
The battery life is 7,000 10-second measurements.
ac
ac
3
2
and 45Hz<frequency ≤65Hz
and frequency<45Hz
1
10 20 30 40 50 60 70 80 90
36
°C

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

C.a 762 ip2x

Table des Matières