Change Invia Motion canister / tubing set
Acoustic signals are muted by pressing the
1
Wash hands and prepare clean paper towel or tissue.
2
Remove the Invia Motion NPWT pump from the carrying case. Switch the Invia Motion NPWT
pump off by pressing and holding the
3
Position the clamp next to the quick-connector
and close the clamp on the pump tubing and
also close the clamp on the dressing tubing.
4
Disconnect the Invia Motion NPWT pump tubing from
the dressing tubing by pressing on the sides of the
Quick-connector.
Use a paper towel or tissue to cover the tip of the used
tubing to protect against wound fluid leakage.
5
Detach the full canister/tubing set
by rotating the canister to the left.
6
Wash hands.
Obtain and open a new Invia Motion canister/tubing set package.
7
CAUTION
Do not use an Invia Motion canister/tubing set if the sterile packaging is damaged
or has expired.
Make sure the correct canister is used: Ref 087.4016.
Contact your healthcare professional if the correct canister is not available.
8
Remove the canister/tubing set from the packaging and
attach it to Invia Motion NPWT pump.
9
Connect the Invia Motion NPWT pump tubing to the
dressing tubing. Push together until you hear a click.
10 Open the clamp on the dressing tubing. Also make sure the clamp on the pump is open.
11 Switch the Invia Motion NPWT pump on by pressing the
pump into the carrying case.
12 Ask your clinician how to properly dispose of a used canister/tubing set.
Medela AG
USA
Lättichstrasse 4b
Medela LLC
6340 Baar, Switzerland
1101 Corporate Drive
McHenry, IL 60050
www.medelahealthcare.com
USA
Phone +1 877 735 1626
Fax
+1 815 307 8942
info-healthcare@medela.com
www.medelahealthcare.us
button.
button for 3 seconds. Press only once.
Clamp open
Clamp closed
button. Slide the Invia Motion NPWT
United Kingdom
Medela UK Ltd.
Huntsman Drive
Northbank Industrial Park
Irlam, Manchester M44 5EG
United Kingdom
Phone +44 161 776 0400
Fax
+44 161 776 0444
info@medela.co.uk
www.medelahealthcare.co.uk
Notification
Symbol
Description
Leak in system
blinking + acoustic
signal
Blockage
blinking + acoustic
signal
Canister full
blinking + acoustic
signal
Battery low
blinking + acoustic
signal
Battery fully discharged
blinking + acoustic
signal
Battery missing
blinking + acoustic
signal
Pump time expired
blinking
(example for 15 days
version)
Error (Internal fault)
acoustic signal
Acoustic Signal can be muted by pressing
CAUTION: U.S. Federal law restricts this device to sale by or on the order of a physician.
Medela, Invia and Invia Motion are registered trademarks of Medela Holding AG, Switzerland.
Canada
Switzerland
Medela Canada Inc.
Medela AG
4160 Sladeview Crescent, Unit #8
Lättichstrasse 7
Mississauga, Ontario, L5L 0A1
6340 Baar
Canada
Switzerland
Phone +1 905 608 7272
Phone +41 848 633 352
Fax
+1 905 608 8720
Fax
+41 41 562 14 42
info@medela.ca
contact@medela.ch
www.medelahealthcare.ca
www.medelahealthcare.ch
6
Troubleshooting
Check the system for leaks – refer to
instructions for use, chapter Notifications and
troubleshooting
Check the system for blockage – refer to
instructions for use, chapter Notifications and
troubleshooting
Change the Invia Motion canister/tubing set –
refer to description in chapter "Change Invia
Motion canister/tubing set"
Connect the Invia Motion pump to the power
supply and fully charge battery. Charging
time can be up to 4 hours – refer to the
instructions for use, chapter Notifications and
troubleshooting
Connect the Invia Motion pump to the power
supply – refer to the instructions for use, chapter
Notifications and troubleshooting
Ensure that battery compartment cover is
mounted and closed properly – refer to the
instructions for use, chapter Notifications and
troubleshooting / Overview notifications
Dispose of the Invia Motion NPWT pump.
Attach another device if necessary – refer to the
instructions for use, chapter Notifications and
troubleshooting
Switch the Invia Motion NPWT pump off
by pressing the
button and holding for
3 seconds.
Patients: contact your healthcare professional
immediately for further instructions. Refer to the
instructions for use, chapter Notifications and
troubleshooting / Overview notifications
International Sales
Medela AG
Lättichstrasse 4b
6340 Baar
Switzerland
Phone +41 41 562 51 51
Fax
+41 41 562 51 00
ism@medela.ch
www.medelahealthcare.com

Publicité

Table des Matières
loading

Table des Matières