Télécharger Imprimer la page

Health O Meter PRO PLUS 2101KL Mode D'emploi page 31

Masquer les pouces Voir aussi pour PRO PLUS 2101KL:

Publicité

Les langues disponibles

Les langues disponibles

Load Cell D
Battery
PC Communication (USB)
Write to RAM
Read from RAM
Table B - Components to test
The different tables on the scale:
There will be two data tables on the scale. The first will store the subject record data (patient-related data
that does not change) and the second will store the data recorded during the readings made.
Subject record data
Fields: ID, TARE, HEIGHT
Readings
Fields: Index (referring to the ID from the Subject Record Table), Weight, Date-time
Sample and explanation of esc protocol
1. When the scale measures the weight of a patient and sends this over the
communications line to the PC, it will look like this:
<SCALE>
<ESC>R<ESC>I1234567890<ESC>W90.98<ESC>H178.7<ESC>T5.7<ESC>B28<ESC>Nm<ESCE
2. When the PC requests the values of the stored weighings, it will look like this:
First the PC will ask the scale to send the data. We do this with the "Get Subject Record Data and
Readings" command.
<SCALE>
<ESC>G<ESC>E
Then the scale will reply with the stored data.
LCD
E3D
BAT
E4L(Bat Low, but still usable) or
E4U (Bat Low and Unstable)
PCC
E05
WRM
E08
RRM
E09
Direction of communication
Direction of communication
Get subject data
EOP
30
<PC>
Reading (Command)
Patient ID = 1234567890
Weight = 90.98
Height = 178.70
TARE = 5.7
BMI = 28
Unit of measure = Metric
EOP
<PC>

Publicité

loading

Ce manuel est également adapté pour:

Pro plus 2500klPro plus 2600kl