Minimum Capture Measurements
When the probe couples with the work piece it will display the current
measurement. When the probe is lifted away it will display the minimum
value of the measurement carried out while the MIN Indicator flashes for
several seconds. If you continue taking measurements while the MIN
Indicator is flashing, the former measurements will continue to take part
in the minimum value capturing. If you carry out measurements after the
MIN Indicator stops flashing the minimum value capturing will re-start.
When the Minimum capture function is ON the LCD will display the MIN
Indicator.
Two Point Calibration
1.
Select two standard samples of the same material to be
measured, among which one has a thickness equal to or slightly
higher than the tested piece, and the thickness of another test
piece is slightly lower than the tested piece
2.
Before carrying out 2-point calibration, turn off the Minimum
Capture Function and Erase the CAL Data in the Memory Manager
3.
Set the 2-point calibration to ON in the System Setup Menu
4.
Press the Mode Button to return to the main display
5.
Press the Enter Button at any time during measurement to enter
the 2-Point CAL
6.
Measure the thinner standard test piece and use the Up and Down
Buttons to adjust the measurement to standard value
7.
Press the Enter Button and the LCD will prompt to measure the
thicker piece
8.
Measure the thicker standard test piece and use the Up and Down
Buttons to adjust the measurement to standard value
9.
Press the Enter Button when the calibration operation is finished
534 Blue
123 Yellow
485 Red
123 Yellow
534 Blue
1-877-849-2127 | info@reedinstruments.com | www.reedinstruments.com
Only if you REALLY need them:
Blue - 100/80/30/5
Yellow - 0/27/100/0
Red - 10/100/100/5
Yellow - 0/27/100/0
Blue - 100/80/30/5
reedinstruments
www
Rich Black -
REED Instruments
20/20/20/100
10
com