AC/DC CURRENT (LESS THAN 2000A)
AC Current is measured by pressing the
clamp trigger
and placing it around a current-carrying wire.
When measuring, care should be taken to
ensure that the clamp
closed with trigger
that the wire passes perpendicularly through
the center of the clamp
arrow markings
To measure current:
1. Rotate the Function Selector switch
the Current A
2. The meter defaults to AC measurement. To
measure DC, press the "SEL/DC ZERO"
button
to toggle between AC and DC
9
modes. The AC or DC icon on the LCD
indicates which mode is selected. Note
"AC" or "DC" on the display.
3. Place the clamp
align the polarity markings
the wire to avoid negative readings. The current measurement will
be shown in the display. The meter will auto-range to display the
measurement in the most appropriate range.
To measure inrush current:
1. Rotate the Function Selector switch
Place the clamp
Display
1
will show "----". Initiate the Inrush event and the inrush
current will be captured and presented in the display. A second press of
"INRUSH"
10
on the display and the meter will wait for the next Inrush event.
NOTE: The Inrush measurement period is 100 milliseconds, with a 20
millisecond sampling rate.
2. Press and hold "INRUSH"
normal measuring mode.
Disconnect test leads when measuring with the clamp.
NOTE: If non-zero values are displayed prior to measuring in DC
current mode, an offset correction may be required to improve
accuracy. With meter in DC current mode, press and hold the
"SEL/DC ZERO" button
DC current ZERO function. Subsequent DC current measurements
automatically subtract the measurement that was initially present as
an offset correction.
10
OPERATING INSTRUCTIONS
to open the clamp
11
is completely
3
fully released, and
11
in line with the
3
.
12
setting.
around the wire. When measuring DC current,
3
around the wire and press "INRUSH"
3
will reset the Inrush measurement, "----" will be presented
10
for more than one second to initiate the
9
3
to
2
on the clamp with the polarity of
16
2
to the Current A
for more than one second to return to
setting.
. The