Installation; Restriction Of Use; Before Using The Zippette Classic; Assembly - VWR Zippette Classic 2,5 ml Mode D'emploi

Distributeur adaptable sur flacon
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 17
kage
Pac
Contents
Zippette Classic bottle top dispenser
Three adaptors 38 mm, 40 mm and 45 mm
PTFE inlet tube

Installation

Restriction of Use

NEVER use the Zippette
Classic
Liquids which are not compatible with PTFE, PVDF, FEP, borosilicate glass, alumina ceramic or those
which may attack platinum-iridium
Hydrofluoric acid
Liquids which contain solid particles
Temperature limits of Zippette

Before Using the Zippette Classic

Check that the instrument has not been damaged in transit.

Assembly

Classic
The Zippette
is packed with the dispense tube attached and the inlet feed tube removed. The 300
mm length of PTFE inlet tube provided should be trimmed to fit your particular reservoir. The bottom end
should be cut diagonally at a length that is close to the bottom of the container when the Zippette
is assembled to the container.
The threaded platform base of the Zippette
recommended squat amber reservoir 1.4 litre with side filler neck, see page 10 for ordering details. The
Classic
assembled Zippette
Classic's threaded platform base only. Removal should also be by means of hand torque applied to this
same base.
Do not operate the piston until the unit is safely and fully mounted on the reservoir bottle.

Specification

Accuracy ≤±0.3% on maximum delivery and a precision ≤0.1%CV using distilled water at 20 °C.
Testing takes place according to DIN EN ISO 8655 using maximum nominal volume of distilled water at
20 °C, with a uniform and smooth dispense action.
with:
Classic
and reagent are 15 °C to 40 °C
Classic
has a 30 mm screw thread for use with the
is screwed onto the reservoir using gentle hand torque applied to the Zippette
|