Calibration Menu
You can choose the automatic or manual calibration mode, in both cases the calibration at pH 7 is automatic.
Automatic calibration:
-
When the display shows the value of the buffer solution, insert the probe into the bottle, press
display will show the countdown of the 60 seconds required to complete the calibration. If the quality of the
alignment is less than 50%, an error appears on the display; press
pump will automatically exit the calibration); if the quality is higher than 50%, the value appears on the display;
pressing
you will be prompted for the pH 4 or 9 buffer solution; at this point the procedure is similar to the
previous calibration.
-
Manual calibration:
When the display shows the value of the buffer solution, insert the probe into the bottle, press
display will show the countdown of the 60 seconds required to complete the calibration. If the quality of the
alignment is less than 50%, an error appears on the display; press
pump will automatically exit the calibration); if the quality is higher than 50%, the value appears on the display;
pressing
on the display will flash the pH 7.00 value; press
to confirm and start the calibration procedure as described previously.
EM00136493
to insert the value of my solution, then press
rev. 1.0 EN
to exit the calibration (after 4 seconds the
to exit the calibration (after 4 seconds the
and the
and the