Electrical Specifications
Accuracy is given as ± (% of reading + counts of least significant digit) at 23°C ± 5°C,
with relative humidity less than 80% and is specified for 1 year after calibration.
Condition of Auto Power On 
With batteries fitted :`
Without batteries :

Auto Power Off

The meter automatically turns off if one of the following conditions are met for approx
10 seconds:
ƒ The Auto Power On condition is not met.
ƒ Both buttons are not pressed.
The meter automatically turns off if one of the following conditions are met for approx
30 seconds:
ƒ The resistance is OL when the meter is in resistor mode.
ƒ The current is < 1.0 A when the meter is in current mode
For > 300 V, Time rating (tr): 30 seconds; Recovery time (rt): 240 seconds
AC Function :
www.megger.com
ƒ > 3.0 V or < -8.0 V between L2 and L1
ƒ Detect AC signal by Single Pole
ƒ Continuity
> | ±35,0 V DC | or > 45.0 V AC
between L2 and L1
ƒ ACV and ACA specifications are AC coupled,
true RMS.
ƒ For non-sinusoidal waveforms, Additional
Accuracy by Crest Factor (C.F.):
Add 1.0% for C.F. 1.0 ~ 2.0
Add 2.5% for C.F. 2.0 ~ 2.5
Add 4.0% for C.F. 2.5 ~ 3.0
ƒ Max. Crest Factor of Input Signal:
3.0 @ 5000 counts
1.5 @ 10000 counts
Specifications
19