Owner's Guide
Only use authorized spare parts and accessories supplied by
.
®
Coltène
Use of parts other than those recommended may dam-
age circuitry and will void product warranty. Repair or disassembly
of this device must be done by authorized personnel only.
®
Coltène
is not responsible for any damage caused by the
following:
• Improper repair or maintenance service performed by
unauthorized personnel or facilities.
• Use of any unauthorized replacement parts or accessories.
•
S.P.E.C. 3
high-output light energy. Do not look
directly at the light emitted from this
curing light. Retinal damage may ocurr.
•
Patient, clinician and assistants should wear
UV orange eye protection when this device
is in use. Retinal damage may occur.
WARNING
•
After use, place the handpiece on the
charger base securely. Failure to properly
secure the handpiece in the charger base
may result in an incomplete battery charge.
•
Light output from this light may be
significantly increased compared to other
polymerization devices; therefore when
using this device determine the curing time
of your material using the enclosed test
CAUTION
block.
•
Do not drop the handpiece or light guide.
Device damage may occur.
•
Avoid electromagnetic radiation generated
from other electrical appliances, as they
may cause the product to malfunction.
®
LED Curing Light produces
EN
•
Take caution when using this product in the
presence of a flammable anesthetic mixture or
an oxidizer like oxygen or nitrous oxide. This
device may act as an ignition source.
•
DO NOT expose soft oral tissues at close
proximity or burns may occur. Maintain a safe
distance between the light guide tip and the
soft tissue.
•
If using the S.P.E.C. 3
the Standard Mode and in close proximity of
the gingival tissue, DO NOT expose tissue for
more than 5 seconds or minor burns may oc-
cur. If longer curing time is required, consider
a dual-cure product (composite or adhesive).
Irradiance output in High Power Mode is 1600
mW/cm².
•
In 3K Mode, DO NOT expose soft tissue for
WARNING
more than 2 seconds or burns may occur. If
a longer cure is needed, consider a dual cure
product (composite or adhesive). Irradiance
output in 3K mode is 3000mW/cm
•
Barrier sleeves are single patient use only.
Barrier sleeve must be used to prevent patient
cross contamination.
•
Persons having a history of photosensitizing
drugs should not be exposed to light from
this light.
•
Adjust curing techniques in accordance with
the increased energy. Pulpal and soft tissue
damage may occur if excessive energy is
applied to a restoration.
Failure to comply with these warnings may
result in injury to personnel.