Coltene SPEC 3 LED Mode D'emploi page 5

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 138
Owner's Guide
Only use authorized spare parts and accessories supplied by
.
®
Coltène
Use of parts other than those recommended may dam-
age circuitry and will void product warranty. Repair or disassembly
of this device must be done by authorized personnel only.
®
Coltène
is not responsible for any damage caused by the
following:
• Improper repair or maintenance service performed by
unauthorized personnel or facilities.
• Use of any unauthorized replacement parts or accessories.
S.P.E.C. 3
high-output light energy. Do not look
directly at the light emitted from this
curing light. Retinal damage may ocurr.
Patient, clinician and assistants should wear
UV orange eye protection when this device
is in use. Retinal damage may occur.
WARNING
After use, place the handpiece on the
charger base securely. Failure to properly
secure the handpiece in the charger base
may result in an incomplete battery charge.
Light output from this light may be
significantly increased compared to other
polymerization devices; therefore when
using this device determine the curing time
of your material using the enclosed test
CAUTION
block.
Do not drop the handpiece or light guide.
Device damage may occur.
Avoid electromagnetic radiation generated
from other electrical appliances, as they
may cause the product to malfunction.
®
LED Curing Light produces
EN
Take caution when using this product in the
presence of a flammable anesthetic mixture or
an oxidizer like oxygen or nitrous oxide. This
device may act as an ignition source.
DO NOT expose soft oral tissues at close
proximity or burns may occur. Maintain a safe
distance between the light guide tip and the
soft tissue.
If using the S.P.E.C. 3
the Standard Mode and in close proximity of
the gingival tissue, DO NOT expose tissue for
more than 5 seconds or minor burns may oc-
cur. If longer curing time is required, consider
a dual-cure product (composite or adhesive).
Irradiance output in High Power Mode is 1600
mW/cm².
In 3K Mode, DO NOT expose soft tissue for
WARNING
more than 2 seconds or burns may occur. If
a longer cure is needed, consider a dual cure
product (composite or adhesive). Irradiance
output in 3K mode is 3000mW/cm
Barrier sleeves are single patient use only.
Barrier sleeve must be used to prevent patient
cross contamination.
Persons having a history of photosensitizing
drugs should not be exposed to light from
this light.
Adjust curing techniques in accordance with
the increased energy. Pulpal and soft tissue
damage may occur if excessive energy is
applied to a restoration.
Failure to comply with these warnings may
result in injury to personnel.
5
LED Curing Light in
®
2.

Publicité

Table des Matières
loading

Table des Matières