Description of procedures
Rinse cycle at 65°C. Then carefully rinse
Dishwasher
all parts with clean water.
Hand wash in warm water. Use a
Hand wash
mild detergent. Then rinse all parts in
clean water and allow to air-dry.
Clean mechanically, e.g. with a brush,
in dilute solution. After disinfection rinse
Disinfection*
all parts thoroughly with distilled water
and dry thoroughly.
Sterilization
Steam sterilization in devices to comply
(alternative to
with EN 285. Temperature 134 °C,
disinfection)
minimum hold time 3 minutes.
*We recommend the disinfectants
GIGASEPT FF, CIDEX OPA and
ANIOSYME DD1. Evidence of tolerance to
these disinfectants has been provided. It is
essential to follow the manufacturer's
instructions.
Note
We recommend that you wear gloves
(e.g. disposable gloves) while cleaning
and disinfecting your mask.
Wash the HEADstrap headgear before
using for the first time, as the colour might
run.
The headgear may only be washed by
hand!
Do not iron the headgear, otherwise the
hook-and-loop closures will no longer
work!
4.3 After cleaning and disinfecting
Dry all parts thoroughly after cleaning and
disinfecting. Check the components for residual
contamination or damage. If necessary, repeat
the cleaning and disinfecting. Replace
damaged parts. Reassemble the mask.

4.4 Multiple patient use

If the mask is used by a different patient, it must
be cleaned and disinfected first.
Disassemble the mask. Disinfect or sterilize all
parts as described in section "4.2 Cleaning,
disinfection, sterilization".
Use a new headgear and where necessary a
new quick release system.
20

5 Functional check

Before use check that everything is working
properly as described below.
vented variants
Always check that the anti-asphyxia valve is
working before each use as described below:
1.Ensure that the valve and membrane are
dry, undamaged and free of secretions. Do
not use the mask if the valve or membrane
is damaged (e.g. through tears and
deformation). In this case replace the anti-
asphyxia valve. (see "7. Scope of supply"
on page 21)
2.Ensure that the valve membrane is not
closing the valve aperture. Only then can
ambient air flow into the mask.
3.Attach the breathing hose (delivered with
the therapy unit) to the elbow of the mask
and to the therapy unit.
4.Switch on the therapy unit. The membrane
should now close the aperture of the anti-
asphyxia valve. Only then can ambient air
flow from the therapy unit into the mask.
5.Do not use the mask if the valve or the
membrane is not working. Replace the anti-
asphyxia valve. (see "7. Scope of supply"
on page 21)
JOYCE Full Face
masks
Carry out a visual check after each cleaning
and disinfecting procedure.
If the mask cushion is damaged and is no longer
airtight, replace it with a new one.
Check the functionality and air-tightness of the
ball-and-socket joint on the elbow. If it no longer
moves easily or is no longer airtight, replace the
elbow and retaining ring.
Parts that are no longer usable can be disposed
of as normal household waste.