Error messages

Error message
The measurement could not be performed. Please
try again.
Connection error.
Blood pressure measurement could not be
performed.
Causes
Hardware auto- check error.
Communication error between the
blood pressure monitor and the
collateral device.
- Cuff is not deflated before measurement.
- Noise interference.
- Excessive user movements.
Countermeasures
The measurement could not be performed.
Please try again. If the problem occurs again,
please contact customer service.
Push the blood pressure monitor button
for 3 seconds to turn it off. Quit the
application, turn you bluetooth connection
off and on. Push the blood pressure monitor
button to turn it on.
Please wait until the cuff is totally deflated
before starting a new measurement. Stay still
during the measurement. Do not speak during
the measurement.