ECG Performance
Clinical performance
The ECG algorithm ability to accurately classify an ECG
into AFib and sinus rhythm was tested in a clinical trial of
115 subjects. Rhythm classification of a 12-lead ECG by a
cardiologist was compared to the rhythm classification of
a simultaneously collect ECG from BPM Core.
Sensitivity (%)
Atrial Fibrillation
94.57
Sensitivity (%)
Normal Sinus
93.76
Rhythm
`
The ECG recording result on the mobile app provides a
detailed display of the result.
Specificity (%)
93.56
Positive
predictivity (%)
93.78