ResMed Quattro Air Mode D'emploi page 9

Masquer les pouces Voir aussi pour Quattro Air:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Daily/After each use:
• Disassemble the mask components according to the disassembly
instructions.
• Thoroughly clean the separated mask components (excluding
headgear), by gently rubbing in warm (approximately 86°F/30°C) water
using mild soap.
• To optimize the mask seal, facial oils should be removed from the
cushion after use.
• Use a soft bristle brush to clean the vent.
• Inspect each component and, if required, repeat washing until visually
clean.
• Rinse all components well with drinking quality water and allow to air
dry out of direct sunlight.
• When all components are dry, reassemble according to the
reassembly instructions.
Weekly:
• Handwash the headgear.
Reprocessing the mask between patients
Only Sleep Lab Mask (SLM) variants of the mask are intended for
multi-patient re-use. When using between patients, these masks must be
reprocessed according to cleaning and disinfection instructions available
on ResMed.com/downloads/masks.
Troubleshooting
Problem/possible cause
Mask is uncomfortable
Headgear straps are too tight.
Mask could be the wrong size.
Solution
The cushion membrane is designed to inflate
against your face to provide a comfortable
seal with low headgear tension. Adjust
straps evenly. Ensure that the headgear
straps are not too tight and that the cushion
is not crushed or creased.
Talk to your clinician to have your face size
checked against fitting template. Note that
sizing across different masks is not always
the same.
English
5

Publicité

Table des Matières
loading

Table des Matières