Inserting a Specimen
Replacing a
Grinding/Polishing Disc
Starting a Preparation Step
14
LaboForce-3
Instruction Manual

Quick Reference

 Lift the thrust pad to make room for the specimen.
 Place the specimen in the specimen holder plate hole and lower
the thrust pad. Make sure that the specimen diameter matches
the specimen holder plate hole.
 Adjust the force by turning the adjustment screw.
 Stop the grinding/polishing machine.
 Unlock the LaboForce-3 handle and let it move into upright
position.
 Change the disc on the turntable.
 Lower LaboForce-3 again and press gently to lock the machine.
 Lower LaboForce-3 and press gently to lock the machine in
position.
 Place 1 to 3 specimens in the specimen holder plate on the
chosen surface.
 Adjust the lubricant dosing level.
 Press START
Perform the preparation step and press STOP
.
LaboPol
 Stop lubricant dosing
Drip Lubricator: Close the valve on the drip lubricator
LaboDoser: Stops automatically after the pre-set time.