Trouble-Shooting - Struers LaboForce-3 Manuel D'instructions

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Error
The specimen mover
does not start when
LaboPol is started.
Specimen holder plate
vibrates
The preparation disc
runs unevenly or stops
The column starts to
turn
Bad or unusual
preparation results
from the same method
12
LaboForce-3
Instruction Manual

4. Trouble-Shooting

Cause
The switch for the specimen mover on the
back of LaboPol is set to "off"
Specimen holder plate unbalanced
Specimen holder plate screws loose
Force too high
Bottom nut of column is loose
The adjustment screw turns by itself, leading
to differing forces
Action
Set the switch to "on"
Replace specimen holder plate
Tighten specimen holder plate
screws
Lower force
Tighten the nut at the base of
column immediately (See Section
INSTALLATION)
Increase friction by turning the
friction screw in the spring
housing
Replace the friction pins if
necessary