3.5 Sensor Installation

Installation Procedure

1. Perform all required menu steps taking note of the sensor model and size designation.
The transmitter then issues the number index and prompts you to press <ENTER> to
finish the sensor install routine. Stop at this point.
Note
Note the number index value displayed in the [Install Sensor] menu. You will use this
index to properly space the sensors. Check to ensure that you have a matched set of
sensors. They both should have the same S/N number but marked with either an "A" or
"B" (e.g., 19256A and 19256B).
2. Prepare the pipe surface area where the sensors will be mounted. Degrease the surface
and remove any grit, corrosion, rust, loose paint, etc.
Note
Before beginning refer to the Reflect Mount sensor installation diagram below.
Figure 3-7
40
Optional: On larger pipes, multiple
lengths of straps can be linked together
to surround pipe
Mounting Strap positioned around
Mounting Frame
Sensor shown in the 9 o'clock position on
pipe
Mounting Frame
Reflect Mount with Mounting Frames and Spacer Bar
Space Bar Platform & Clamping Screw
Space Bar (Front View)
Metal Post
Mounting Strap Adjusting Screw
FUG1010 IP65 NEMA 4X Quick Start
Operating Instructions, 06/2014, A5E02951527-AC